Best Suitable for 10kΩ~200GΩ chip, melf, lead type of sorting machine of ultra-high resistance, taping and triming machine
?
● Trimming mode is possible.
● High speed integral few method of an error to noise. [Available to make the integral time setting for each range.]
● % measurement is possible by digital setting of the standard resistance value 4 digit.
[±50.00%]
● Comparator result is possible open collector output, LED display and buzzer sound.
● RS-232C interface are built-in as standard equipment.
● Switchable the measuring voltage H/L corresponds to the measument objects
Specifications
〔HV-deviation value measurement〕
| Range | Measuring range | Measuring voltage | Measuring Accuracy |
|---|---|---|---|
| 100kΩ | 5.00kΩ~150.00kΩ | 15V | ±0.03%±1digit |
| 1MΩ | 50.0kΩ~1500.0kΩ | 15V | |
| 10MΩ | 0.500MΩ~15.000MΩ | 15V | |
| 100MΩ | 5.00MΩ~150.00MΩ | 100V | ±0.05%±1digit |
| 1GΩ | 50.0MΩ~1500.0MΩ | 100V | |
| 10GΩ | 0.500GΩ~15.000GΩ | 100V | ±0.2%±1digit |
| 100GΩ | 5.00GΩ~150.00GΩ | 200V | ±0.5%±2digit |
|
200GΩ
(1000GΩ) |
50.00GΩ~300.00GΩ
(300.01GΩ~1500.0GΩ) |
200V | ±1%±4digit |
| ※1 |
〔HV-absolute value measurement〕
| Range | Measuring range | Measuring voltage | Measuring Accuracy |
|---|---|---|---|
| 100kΩ | 5.00kΩ~150.00kΩ | 15V | ±0.1% of rdg ±1digit |
| 1MΩ | 50.0kΩ~1500.0kΩ | 15V | |
| 10MΩ | 0.500MΩ~15.000MΩ | 15V | |
| 100MΩ | 5.00MΩ~150.00MΩ | 100V | ±0.2% of rdg ±1digit |
| 1GΩ | 50.0MΩ~1500.0MΩ | 100V | |
| 10GΩ | 0.500GΩ~15.000GΩ | 100V | ±0.5% of rdg ±1digit |
| 100GΩ | 5.00GΩ~150.00GΩ | 200V | ±0.8% of rdg ±2digit |
|
200GΩ
(1000GΩ) |
50.00GΩ~300.00GΩ
(300.01GΩ~1500.0GΩ) |
200V | ±1.5% of rdg ±4digit |
| ※1 |
〔LV-deviation value measurement〕
| Range | Measuring range | Measuring voltage | Measuring Accuracy |
|---|---|---|---|
| 100kΩ | 5.00kΩ~150.00kΩ | 15V | ±0.03%±1digit |
| 1MΩ | 50.0kΩ~1500.0kΩ | ||
| 10MΩ | 0.500MΩ~15.000MΩ | ||
| 100MΩ | 5.00MΩ~150.00MΩ | ±0.3%±1digit | |
| 1GΩ | 50.0MΩ~1500.0MΩ | ||
| 10GΩ | 0.500GΩ~15.000GΩ | ±1.2%±2digit | |
| 100GΩ | 5.00GΩ~150.00GΩ | ±1.5%±2digit | |
|
200GΩ
(1000GΩ) |
NOT available |
〔LV-absolute value measurement〕
| Range | Measuring range | Measuring voltage | Measuring Accuracy |
|---|---|---|---|
| 100kΩ | 5.00kΩ~150.00kΩ | 15V | ±0.1% of rdg ±1digit |
| 1MΩ | 50.0kΩ~1500.0kΩ | ||
| 10MΩ | 0.500MΩ~15.000MΩ | ||
| 100MΩ | 5.00MΩ~150.00MΩ | ±1.2% of rdg ±1digit | |
| 1GΩ | 50.0MΩ~1500.0MΩ | ||
| 10GΩ | 0.500GΩ~15.000GΩ | ±1.5% of rdg ±1digit | |
| 100GΩ | 5.00GΩ~150.00GΩ | ±2.0% of rdg ±2digit | |
|
200GΩ
(1000GΩ) |
NOT available |
※Range more than 10GΩ is measuring accuracy when appropriate to set measuring time etc.
※1 The accuracy is out of guarantee
※The above accuracy is the value fully shielded
| Indication range | ±50.00% | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Measuring method | 2 terminal measurement | |||||||||
| Measuring time |
|
|||||||||
| Measuring delay time | 0 msec~99msec setting possible every range | |||||||||
| Comparator | Both upper and lower limit:±0.00%~±50.00% LO,GO,HI judgment indication + output | |||||||||
| Control signal |
|
|||||||||
| Warm up time | More than 30 minutes | |||||||||
| Others | Zero, Full scale calibration function by panel face operation RS-232C interface are built-in as standard equipment. | |||||||||
| Power supply | AC100V~AC240V, Automatic switching formula, 50/60Hz | |||||||||
| Outer dimension | 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) | |||||||||
| Weight | about 4kg |
The Outline
AX-1644B, digital ultra-high resistance checker, can be measured from 10kΩ to 1000GΩresistance value, high accuracy, high stability, and fast speed by digital measurement, and outputs to judge LO/GO/HI decision built in digital comparator.